49
Print graph Option "Yes": A graph will be printed with the diagram.
Option "No": No graph will be printed. The preselected option
is "Yes".
Search by Select between "customer name" and "customer number".
Depending on this selection, the name or the number of the
customer can be selected in the customer menu.
Pitot factor The pitot factor is important for the measurement of the flow
speed. It depends on the cross section of the utilized s-tube.
The pitot factor of the Wöhler S-tube probe is 0.93. Therefore
the preselected value is 0.93.
If a “normal” S-tube is connected, the pitot factor 1.0 has to
be entered.
Thermal capacity The default setting of the specific heat capacity is 0,2790
Wh/kg °C. The value is important for the measurement of the
ventilation loss (heating check)
Reference temperature The reference temperature can be adjusted from - 30 °C to +
70 °C.
Alpha Heat transmission coefficient for the u-value measurement
Language Depending on the version, the languages English, German or
other languages can be selected.
Printer logo Enter a custom corporate logo (6 lines) that will appear in
every printout.
Default Reset the PC 410 to its factory default setting (except the
calibration).
6.13 Option "Save measurements" To save all records which are marked by a tick mark in the
main menu, select the "Save measurement" option.