8. TECHNICAL DATA
Measuring ranges
Table 13
ErCAo
Lack of calibration of analog outputs. Press any button
to transit to the normal work. Analog outputs will not be
serviced. Contact the service workshop.
ErCAL
Error of calibration. The work is stopped – The meter is
not in the state to carry out measurements in a correct
way. Incorrect checksum of calibration coefficients or
lack of calibration.
Kind of input Indication range Class
Pt100
-205...855°C (-200...850°C)
0.1
Pt500
Pt1000
400 W 0...410 W (0...400 W)
4000 W 0...4010 W (0...4000 W)
Thermocouple of J type -220...1210°C (-200...1200°C)
Thermocouple of K type -280...1382°C (-270...1370°C)
Thermocouple of N type -250...1310°C (-240...1300°C)
Thermocouple of E type -280...1010°C (-270...1000°C)
Thermocouple of R type -55...1775°C (-50...1770°C)
Thermocouple of S type -55...1775°C (- 50...1770°C)
Voltage input 0...10 V -13...13 V (-10...10 V)
Current imput -24...24 mA (- 20...20 mA)
Voltage input 60 mV -10...63 mV (0...60 mV)
Current time 00.00...23.59 0.5 sec/24 h
Additional error of thermocouple cold junction temperature compensation: 0.1%
of measuring range
Additional error of wire resistance compensation: 0.1% of the 400 W range