13
fig. 10-5
5. Pressbutton“M1”or“M2”torecordthemeasurementin“M1”or“M2”memory,respectively.
Forexample,ifbutton“M2”ispressed,thedisplaywillalternateasshowninFig.10-5.Ifneither
button“M1”nor“M2”ispressed,themeasurementwillnotberecorded.
6. Pressbutton“O/|”toreturntostandbymode.Waitforatleast5minutesbeforetakinganother
measurement. If the device remains unused for 3 minutes, it will return to standby mode
automatically.
Automatic Inflation
This device has 4 levels of inflation pressure: 190 mmHg, 230 mmHg, 270 mmHg and 300 mmHg. When
190 mmHg is insufficient to obtain accurate measurement, or arm movement occurs, the device
will automatically inflate to an appropriate pressure level in order to obtain the blood pressure
measurement.
Rapid Deflation During Measurement
If you do not feel well during measurement, or want to stop measurement for any reason, press
button“O/|”toquicklydeflatethecuffandreturntostandbymode.