Aethalometer
Model AE33
____________________________________________________________________________________________________________
User's manual Ver. 1.53 July 2015 15/75
1.3 Basic calculations
Determination of black carbon concentration is based on measurement of light absorption on
filter loaded with aerosols:
Optical attenuation:   
󰇛
󰇜
I
0
=reference signal;
I=spot signal
Flow:


󰇛
󰇜

=measured flow
=leakage factor
Attenuation coefficient:


󰇛
 
󰇜


S=spot size; t=time
Absorption coefficient:


C=multiple scattering parameter
(Weingartner et al. 2003)
Black carbon concentration: 



=mass absorption crossection
Loading effect compensation:  

󰇛

󰇜
k=compensation parameter
Final equation: 

󰇛


󰇜
󰇛

󰇜



󰇛

󰇜

Reference: I
0
Sensing I