User's manual Ver. 1.53 July 2015 73/75
7.2 Instrument status
Normal operation
Warning ; Instrument is still
performing measurements, but
there is/was an issue, that needs
to be checked
Instrument stopped. Immediate
response needed.
Instrument status number and status image are presented on the home screen. Pressing on the
status image opens status deconvolution page.
Status
codes
Description
Problem
Solution
4,8,12
Flow status
flow not calibrated
calibrate flow
clogged inlet line
check inlet lines, inlets
32, 48
LED status
Led calibration error
remove flow calibration pad
update firmware
faulty LED board -> service
needed
64
Chamber
status
Chamber locked
Unlock optical chamber
128, 256
Filter tape
status
Not much filter tape
remaining
Schedule filter tape replacement
384
Filter tape
status
No filter remaining
Replace filter tape
Filter not moving
Tighten right spool nut
8192
External
Connection with the
Check physical connection