5.2 Instrument Status
Status of the instrument is currently a status flag (16bit value) consisting of detector (bit 0,1),
flow (2,3), LED (4,5), tape advance(7,8) flags, and will be latr extended to chamber (6), valve
(9), settings (10) , and other important status notifications.
Tape advance (tape advance, fast
calibration, warm-up)
First measurement – obtaining ATN0
Flow low/high by more than 0.25 LPM or
F1 < 0 or F2/F1 outside 0.2 – 0.75 range
Check flow status history
Flow low/high & check flow status
history
Calibration error (at least one channel
OK)
LED error (all channels calibration error,
COM error)