____________________________________________________________________________________________________________
User's manual – Ver. 1.53 July 2015 57/75
5.6.8 Datalogger Qair protocol
The instrument supports data acquisition over protocol which complies with “Communication
protocol for digital links with air quality data acquisition stations ver. 3.1”. It is a limited
version of the industrial JBUS communication protocol.
Standard command format: <slave number><function code><address><number of
words><CRC16>
<Slave number> - Instrument External ID, values from 1 – 255. (1 byte)
<Function code> - Supported function code is 3. (1 byte)
<address> - Supported address values are 0,10,20,30,34,37,179,187 (all decimal). (2 bytes).
<Number of words> - Number of words in the response message. (2 bytes)
<CRC 16> - checksum. (2 bytes).
Examples:
Manufacturer (address code 0)
o Command: [01] [03] [00] [00] [00] [0A] [C5] [CD]
o Response: [1] [3] [14] [4D] [61] [67] [65] [65] [20] [73] [63] [69] [65] [6E] [74] [69]
[66] [69] [63] [20] [20] [20] [20] [5C] [8E]
Serial number (address code 10)
o Command: [01] [03] [00] [0A] [00] [0A] [E5] [CF]
o Response: [1] [3] [14] [41] [45] [33] [33] [2D] [53] [30] [30] [2D] [30] [30] [30] [32]
[37] [20] [20] [20] [20] [20] [20] [62] [9E]
Software and firmware version (address code 20)
o Command: [01] [03] [00] [14] [00] [0A] [85] [C9]
o Response: [1] [3] [14] [31] [2E] [31] [2E] [37] [2E] [31] [20] [35] [31] [38] [20] [20]
[20] [20] [20] [20] [20] [20] [20] [F6] [E]
Date (address code 30)
o Command: [01] [03] [00] [1E] [00] [04] [24] [0F]
o Response: [1] [3] [8] [30] [32] [30] [37] [32] [30] [31] [35] [7D] [A3]
Time (address code 34)
o Command: [01] [03] [00] [22] [00] [03] [A5] [C1]
o Response: [1] [3] [6] [31] [30] [34] [38] [35] [30] [81] [1E]
Operating parameters (address code 37)