Aethalometer
Model AE33
____________________________________________________________________________________________________________
User's manual Ver. 1.53 July 2015 55/75
Supported devices:
o AMES_TPR159 (ID_com = 1) dedicated ambient temperature + relative humidity +
pressure sensor used for flow reporting at ambient conditions
o Comet_T0310 (ID_com = 2) temperature sensor
o Vaisala_GMP343 (ID_com = 3) CO2 sensor
o TSI_4100 (ID_com = 4) mass flowmeter
o Datalogger_AE33_protocol (ID_com = 5) the selected port is waiting for
communication with the data-logger
o Aerosol_inlet_dryer (ID_com = 6) temperature, RH, dew point sensor
o Datalogger_BH_protocol (ID_com = 7) Bayern - Hessen protocol
o Datalogger_Qair_protocol (ID_com = 8) communication protocol for digital links with
air quality data acquisition stations
5.6.1 AMES_TPR159
Data received from AMES is stored at the end of AE33 dataline in the fields_i section in
following format: Temperature, RH, Pressure. If ambient flow reporting standard is selected,
the values Temperature and Pressure in dataline are the same as in the fields_i section.
5.6.2 Comet_T0310
Comet sensor outputs temperature. Data is stored in the fields_i section.
5.6.3 Vaisala_GMP343
Data received from Vaisala is stored at the end of AE33 dataline in the fields_i section in
following format: CO
2
, CO
2
RAW, Temperature.
5.6.4 TSI_4100
TSI is a mass flowmeter. It is used for Automatic calibration, in this case it has to be connected
to COM3. Data received from TSI is Flow. Data is stored in the fields_i section.