Aethalometer
Model AE33
____________________________________________________________________________________________________________
User's manual Ver. 1.53 July 2015 72/75
7. TROUBLESHOOTING GUIDE
7.1 Startup screen checks
Check
Description
Error
Solution
Communication
communication PC to
optical chamber controller
hardware
problem
check cables
Instrument
data
Obtain data (serial number)
from the optical chamber
controller
hardware
problem
check cables
Storage
CF card operation
CF card error
get new CF card 6 SW
Configuration
settings
read setting from the setup file
Setup file error
restore setup file from one of
the older setup files
Valves
operation of the ball valve
ball valve not
moving
check cables
Chamber
optical chamber movement
test
locked chamber
unlock chamber
hardware error
service needed
Pump & Flow
test if pump is working
pump
service needed
tube connections
reconnect tubes
Device
monitoring
Win CE operating system
test
faulty
application file
get new CF card & SW