EasyManua.ls Logo

MAGEE Scientific AE33 - Serial Commands for Communication with the Aethalometer

Default Icon
75 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Aethalometer
Model AE33
____________________________________________________________________________________________________________
User's manual Ver. 1.53 July 2015 53/75
5.5 Serial Commands for Communication with the Aethalometer
All commands finish with a Carriage Return.
$AE33:Dnnn
nnn number of lines
of data requested
Request nnn last
measurements
$AE33:TyyyyMMddHHmmss
yyyy
MM
dd
HH
mm
ss
year
month
day
hour (24)
min
second
Synchronize time with
the data-logger
$AE33:A
Tape advance left
$AE33:SG
Get setup file
$AE33:SSn m
Set setup parameter
n parameter value
m parameter value
Example:
$AE33:SSTimebase 1
Note: instrument must
be stopped
$AE33:L
Get log
Instrument sends last
120 log entries
$AE33:Fn m
Get file(s) starts
streaming number of
files
n folder name (Data,
NDtest, CleanAirTest,
StabilityTest,
InletLeakageTest,
LeakageTest, Flow)
m number of files (0
all)
$AE33:E
End stream