Aethalometer
Model AE33
____________________________________________________________________________________________________________
User's manual Ver. 1.53 July 2015 45/75
01
128
Tape warning (less than 30 spots left)
10
256
Tape last warning (less than 5 spots left)
11
384
Tape error (tape not moving, end of tape)
Tests &
procedures
12, 11, 10
000
No test
001
1024
Stability test
010
2048
Clean air test
011
3072
Change tape procedure
100
4096
Optical test
110
6144
Leakage test
External
device
13
0
0
Connection OK
1
8192
Connection Error
8
7
6
5
4
3
2
1
0
0/1
0/1
0
0/1
0/1
0/1
0/1
0/1
0/1
Table 1. Status flags and explanations.
When a status flag 3 (decimal) is encountered, the Aethalometer stops. In all other statuses, it
continues to operate with a warning, and the data is flagged accordingly.
IMPORTANT!
Valid data is all data where STATUS does not contain numbers 1, 2, 3, 1024, 2048, 3072, 4096,
6144.