Aethalometer
Model AE33
____________________________________________________________________________________________________________
User's manual Ver. 1.53 July 2015 46/75
The overall status is a sum of the statuses in the Table 1. These status flag add, for example:
Status = 0, all OK -> front panel LED's GREEN
0
0
0
0
0
0
0
0
0
Status = 1, all OK, tape advancing -> blinking GREEN LED
0
0
0
0
0
0
0
0
1
Status = 128, machine is running, tape advance warning flag is set -> YELLOW
0
1
0
0
0
0
0
0
0
Status = 289, 289 =256 + 32+ 1 -> machine is doing tape advance, in led calibration 1 or more
channels (but not all) had errors and only 5 or less TA left
1
0
0
1
0
0
0
0
1
Deconvolution of instrument status is triggered by pressing the status color icon on home screen.