EasyManuals Logo

Maico MA 41 User Manual

Maico MA 41
73 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #14 background imageLoading...
Page #14 background image
Operation Manual MA 41
8100456 Rev.9
10
01/16
4.2 Calibration of the Device
The instrument, headphones, bone conduction oscillator, as well as the
optional insert phones and speakers, will come calibrated to the instrument and
have the same serial number (e.g. MA1234567). Use of transducers not
calibrated to this particular instrument will likely lead to incorrect thresholds
causing incorrect test results which will invalidate the test. If a transducer needs
to be replaced, the instrument must be recalibrated with the new transducer.
The use of non-calibrated audiometers can lead to incorrect measurements!
4.3 Where to Setup
The MA 41 should be operated in a quiet room, so that the audiometric
examinations are not influenced by outside noises. Ambient sound pressure
levels in an audiometric test room shall not exceed the values specified in the
norm ISO 8253-1:2010 or ANSI S3.1-1999. For use in noisier environments,
headphones with optional sound insulation muffs are available.
Electro-medical instruments, which emit strong electromagnetic fields (e.g.
microwaves or radiotherapy devices), can influence the function of the
audiometer. Therefore, it is not recommended to the use these instruments in
close proximity to the audiometer as it may lead to incorrect test results.
The test room must be at a normal temperature, usually from 15° C / 59° F to
35° C / 95° F, and the instrument should be switched on approximately 10
minutes before the first measurement. If the device has been cooled down (e.g.
during transport), please wait until it has warmed to room temperature before
using.
External devices such as a computer, printer or
Ethernet which are connected to the device must meet
electrical safety requirements, such as IEC 60601-1 or
UL 60601-1. This is to avoid electrical shock to the user
or the patient. See chapter 3 for more information
WARNING

Other manuals for Maico MA 41

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Maico MA 41 and is the answer not in the manual?

Maico MA 41 Specifications

General IconGeneral
BrandMaico
ModelMA 41
CategoryMedical Equipment
LanguageEnglish

Related product manuals