EasyManua.ls Logo

Maico MA 41 - Page 67

Maico MA 41
73 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation Manual MA 41
8100456 Rev.9
63
01/16
Calibration values: Sound field (0 degree incidence)
Reference equivalent threshold sound pressure level and maximum hearing
levels
For Canton CD 220 / MAICO SBC speaker
ISO 389 7 and ANSI S3.6-1996
Frequency
[Hz]
Tone
RETSPL
[dB
SPL
]
NBN
RETSPL
[dB
SPL
]
Max level
Tone [dB
HL
]
CD220 / SBC
Max level
NBN [dB
HL
]
CD220 / SBC
125
22.0
22.0
65 / -
55 / -
250
11.0
11.0
75 / -
70 / -
500
4.0
4.0
85 / 85
75 / 75
750
2.0
2.0
85 / 85
75 / 75
1000
2.0
2.0
85 / 85
75 / 75
1500
0.5
0.5
90 / 85
80 / 75
2000
-1.5
-1.5
90 / 85
85 / 75
3000
-6.0
-6.0
90 / 90
85 / 80
4000
-6.5
-6.5
95 / 90
85 / 80
6000
2.5
2.5
85 / 90
80 / 80
8000
11.5
11.5
80 / 75
70 / 65
Signal
IEC60645-2
RETSPL
[dB
SPL
]
IEC Max
Level [dB
HL
]
CD220 /
SBC
ANSI S3.6
RETSPL
[dB
HL
]
ANSI Max
Level [dB
HL
]
CD220 /
SBC
Speech
0.0
90 / 80
14.5
90 / 80
SN
0.0
90 / 75
14.5
90 / 75
WN
0.0
90 / 75
0.0
90 / 75

Other manuals for Maico MA 41