PROGRAMMING SECTION MODEL YB8
April 1999/Manning Environmental Inc.
3DJH%
Sampler Configuration Functions
There are 3 major configuration functions the user must be concerned with (*99, *20, & *19). Step-by-step
programming instructions and descriptions o f each function are detailed on the pages that follow :
*99 Sampler Set-Up
*99 allows the user to set the sampler’s configuration. For proper operation, it is critical the unit is correctly
configured. The memory comes preset with the sampler’s defaults. These defaults can be reviewed or
changed by entering the configuration mode (explanations and step-by-step instructions are given below).
Once entries have been made in *99, re-entering the configuration mode is not necessary unless changes in the
data are needed.
Display on LCD Explanation
SAMPLER READY
04/30 04:30:02
This display shows t he sampler is ready t o program. I t displays the
current time. From here the user can enter any TIME, FLOW, or *
Mode. Press the * key to access the * Mode.
ENTER * MODE?
__ __ 04:30:02
Prompts the user to enter either a program or the configuration
function. Press 99 and <ENTER> to configure the sampler.
SAMPLER SETTING?
__ 04:30:02
Sets the sampler to a specific type of operation or bottle
configuration:
1 = Single Bottle
2 = Multi-Bottle
Other numbers are not valid and will cause t he sampler to
malfunction. Enter the desired configuration and press <ENTER>
# OF BOTTLES?
__ __ 04:30:02
Sets the number of bottles (1,2,3,4,6,8,12, or 24) in the sampler. If 1
was selected for the SAMPLER SETTING above, this prompt is
bypassed. Input the number of bottles and press <ENTER>.
SAMPLE VOLUME?
__ __ __ __ 04:30:02
Sets the sample volume (in milliliters), to be collected per sample.
TUBE LENGTH?
__ __ __ __ 04:30:02
Records the length (1 to 99 feet) of the sample intake tube.
DRAW HEIGHT?
__ __ __ __ 04:30:02
Sets the draw height o r lift height (1 to 25 feet) o f t he sampler.