MODEL YB8 PROGRAMMING SECTION
April 1999/Manning Environmental Inc.
3DJH%
Flow Mode - Pulse Accumulation
FLOW Mode - Pulse Accumulation operates the same as FLOW Mode except instead of taking a sample after
every contact closure, a sample is taken after a set number of contact closures (from 2 - 9,999) have been
accumulated. NOTE: This program uses <DELAY START> for setting the number of contact closures to be
accumulated. T he display will show t he number of contact closures the sampler is programmed t o accumulate
before taking a sample. Every time a contact closure is received, the sampler will decrease the number needed
on the display by one. This shows how many more contacts have yet to be accumulated before a sample is
taken. Once the set number of contact closures are received, the sampler will advance t he spout, draw 1
sample and then place it in a bottle. It will t hen wait for t he next accumulation. This will continue until t he
total number of bottles the unit is configured for (set in *99) each have a sample placed in them or a bottle full
condition occurs. If, either of these two conditions occur, the sampler ends the program.
Display on LCD Explanation
SAMPLER READY
04/30 04:30:02
This display indicates the sampler is ready t o program and displays
the current t ime. P ress <FLOW> t o begin programming.
PUSH START/OPTIONS
04:30:02
To set the number of contacts to be accumulated in FLOW Mode -
Pulse Accumulation, press <DELAY START> and then the
<START> button.
DELAY IN PULSES?
__ __ __ __ 04:30:02
The user is now prompted to set the number of contact closures the
sampler will accumulate before taking a sample (2 - 9,999). Until it
is changed or ends it s cycle, it will always accumulate the same
number of pulses before taking a sample.
PUSH START/OPTIONS
04:30:02
Unless add-on options to the program are desired, press <START>.
FLOW MODE
__ __ __ __ 04:30:02
This display shows the number of contact closures remaining before a
sample will be taken. As contact closures are received the sampler
counts down until it reaches 0. It will t hen take a sample and reset to
accumulate the entered number of contact closures again.