Operating and
Instruction Manual
Manual No. YB
Revision 0-600 w/12 March 2014 Errata
MODEL YB8
STATIONARY
PERISTALTIC SAMPLERS
Manning Environmental, Inc.
101 Bar T Drive
Florence, Texas 76527-4445 USA
Office: (800) 863-9337
(254) 793-9955
Fax: (254) 793-9965
www.manningenvironmental.com

Table of Contents

Questions and Answers

  • S
    stevenramosAug 18, 2025
    Why is my Manning Controller system non-responsive?
    • G
      Gregory JohnsonAug 18, 2025
      A non-responsive Manning Controller system can result from several causes: * A tripped circuit breaker: Turn the switch off and then back on. * A loose connection: Check and tighten connectors on the circuit board if necessary. * Controller lock-up: Push the hard RESET button on the lower left side of the processor board. * Controller failure: In this case, remove and replace the controller.
  • R
    roger13Aug 21, 2025
    What to do if Manning Controller purges constantly?
    • B
      brandon17Aug 21, 2025
      If your Manning Controller is constantly purging, the issue could stem from a controller failure, in which case you should remove and replace the controller or the failed board. Alternatively, it might be a motor relay failure; check to see that the relay switches from purge to draw mode.
  • C
    Crystal ThompsonAug 23, 2025
    Why is my Manning Controller working inconsistently?
    • L
      Lawrence GarciaAug 23, 2025
      Inconsistent performance of your Manning Controller can be due to faulty wiring, so check the wiring, starting with the power connections. Another potential cause is a failing controller, which would require you to remove and replace the controller or the failed board.
  • R
    rwallaceAug 26, 2025
    What to do if Manning Controller pump operates but there is no fluid?
    • D
      Desiree LozanoAug 26, 2025
      If the pump in your Manning Controller operates but no fluid is being moved, the problem could be a clogged hose or line, which you should flush with water to clear. Another possibility is a failing peristaltic pump; check its operation, ensuring the drive belt is intact, the drive pulleys and rollers are rotating freely, there is no impediment in the pump tubing, and the motor is humming.
  • B
    benitezmarkAug 27, 2025
    How to fix weak draw in Manning Controller?
    • N
      Nicole SweeneyAug 27, 2025
      A weak draw in your Manning Controller could be caused by several factors: * A pinched intake hose: Check the hose for pinches or damage, and replace it if damaged. * A clogged hose or line: Flush with water to clear any clogs. * An air leak: Inspect the pump tubing for damage. * A failing peristaltic pump: Verify proper operation, ensuring the drive belt is intact, drive pulleys and rollers rotate freely, there are no impediments in the pump tubing, and the motor is humming.
  • S
    simmonschristopherAug 30, 2025
    What causes low sample volume in Manning Controller?
    • N
      Nathan RobertsAug 30, 2025
      Low sample volume in your Manning Controller can be attributed to a few potential issues: a malfunctioning fluid detector, which requires checking the calibration of the fluid sensor or replacing it; excessive tube wear, necessitating replacement of the pump tubing; or a malfunctioning RPM counter, which would require replacing the RPM counter.
  • C
    Christopher WebbSep 1, 2025
    What causes excessive sample volume in Manning Controller?
    • J
      Jennifer GallagherSep 1, 2025
      Excessive sample volume in your Manning Controller might be due to a malfunctioning fluid detector, in which case you should check the calibration of the fluid sensor or replace it. It could also be caused by a malfunctioning RPM counter, requiring you to replace the RPM counter.
  • R
    robertleeSep 4, 2025
    What to do if Manning Controller pump rotor does not rotate?
    • Z
      Zachary HoodSep 4, 2025
      If the pump rotor on your Manning Controller does not rotate, check the pump for proper operation. Ensure that the drive belt is intact, the drive pulleys are rotating, the rollers are rotating freely, there is no impediment in the pump tubing, and the motor is humming.
  • H
    Helen RobertsSep 6, 2025
    What to do if Manning Controller does not respond to command?
    • S
      smithcrystalSep 6, 2025
      If your Manning Controller does not respond to commands, it may be due to an active password. Enter the password when prompted.
  • R
    Rachel PaceSep 8, 2025
    What to do if Manning Controller shows blank display?
    • P
      Pamela HolmesSep 8, 2025
      If your Manning Controller's display is blank, first check to ensure the sampler has power. If power is not the issue, it could be a display failure, in which case you should check the connections and possibly replace the display.

Summary

Manning Environmental Limited Factory Warranty

Warranty Limitations

Exclusions from warranty coverage, including damage from natural disaster, misuse, or unauthorized repair.

Installation and Operation

Introduction

Overview of the YB8 sampler, its features, and manufacturer's commitment to quality.

Hardware and Components

Details on sampler parts, specifications, and subassemblies.

Assembly, Installation, and Operation

Steps for physical assembly, installation, power connection, basic operation, and external connections.

Appendices

Appendix A: Returning Equipment

Procedures for equipment repair and return, including RMA and shipping instructions.

Programming

Introduction and Configuration

Overview of programming modes and initial sampler setup.

Core Configuration Functions

Key setup parameters like *99, calibration, pump utilities, and data logging.

Sampling Program Modes

Various sampling program types including General, Multi-Bottle, Analog, and Composite.

Maintenance

Suggested Maintenance Schedule

Routine upkeep tasks for the sampler, including frequency and time estimates.

Peristaltic Pump Maintenance

Specific upkeep procedures for the peristaltic pump, including tubing replacement.

Troubleshooting

Diagnosing and resolving common problems with the sampler.

Manning YB8 Specifications

General IconGeneral
BrandManning
ModelYB8
CategoryController
LanguageEnglish