MODEL YB8 PROGRAMMING SECTION
April 1999/Manning Environmental Inc.
3DJH%
*07 Flow Mode - Time Interval Delay
*07 operates much like basic FLOW mode except a time interval delay is added after a contact closure has
been received. Just like in FLOW mode, the sampler waits for receipt of a contact closure. Once that closure
has been t aken, the unit immediately begins counting down a user set t ime interval delay. O nce t he delay has
counted down, the unit performs a sampling sequence. I t will then wait fo r the next contact closure while
displaying a running tally indicating the number of samples collected to that point. Upon the next closure the
unit will o nce again count down t he user set interval and then t ake a sample. T his will continue until the total
number of bottles the unit is configured for (set in *99) each have their set number of samples placed in them
or a bottle full condition occurs. If either of these two conditions occur, the sampler ends the program.
Display on LCD Explanation
SAMPLER READY
04/30 04:30:02
This display indicates the sampler is ready t o program and displays
the current time. Press the * key t o begin programming.
ENTER * MODE
__ __ 04:30:02
The sampler is now prompting fo r a star mode to be input. E nter the
numbers which represent the star mode of choice, in this example 02,
and press <ENTER>.
TIME DELAY?
__ __ : __ __ 04:30:02
Input the time the sampler is to wait, after receipt of a contact
closure, to take a sample. A fter getting t he signal the unit will count
down the interval and take a sample. Press <ENTER> after input.
PUSH START/OPTIONS
04:30:02
If no add-on options are desired, press <START> to begin the
program.
FLOW MODE (* 07)
__ __ __ __ 04:30:02
The sampler is now ready to receive contact closures.