MAINTENANCE MODEL YB8
April 1999/Manning Environmental Inc.
Page C-8
water to flush any material that was removed from the liquid sensor out of the sample
tract.
Bottle Full Sensor (Single Bottle Units Only)
1. Locate the bottle full sensor in the neck of the bottle. Either remove the leads from the
ends of the bottle full sensor probes or disconnect the black rubber mating connector.
Remove the bottle full sensor from the discharge
line by sliding it o ff.
2. The unit is made of high grade stainless steel which
is approved by the EPA for toxic or non-toxic
applications. Clean the probes, ensuring that their
is no buildup on the stainless steel probes.
3. Reconnect the unit, reversing the procedure used to remove it. If you disconnected the
leads from the stainless steel probes, make sure they are firmly attached.
Spout (Multiple Bottle Units Only)
1. Remove the spout from the upper union by gently pulling the spout gear from below.
(Rotating back and forth may help).
2. Wash the spout with the appropriate cleaning
solution. Use a test tube brush to clean the internal
surfaces of the spout and upper union.
3. Rinse thoroughly with clean water (warm water is
best) and re-assemble. When replacing the spout,
the timing mark o n t he stepper motor gear must align with t he t iming mark o n t he spout
gear.
Sample Containers
1. Wash with the appropriate cleaning solution. Use a test tube brush to clean the internal
surfaces.
2. Rinse thoroughly in clean water (warm water is
best).
3. Autoclave glass bottles, if desired. Do not
autoclave plastic bottles or caps
since they are constructed of polyethylene.
Refrigerator