INSTALLATION AND OPERATION MODEL YB8
April 1999/Manning Environmental Inc.
Page A-2
4. Program the YB8.
Hardware
Functional Specifications:
Size Control Unit: 10.75" (27.305cm)W x 7" (17.78cm)H x 9.75" (24.765cm)D
Refrigeration Unit: 23.875" (60.65cm)W x 34.5" (87.63cm)H x 24" (60.96cm)D
Total Unit: 23.875" (60.65cm)W x 41.5" (105.41cm)H x 24" (60.96cm)D
Weight Dry Weight: 110 lb (49.89 kg) with refrigerator.
Environmental Nema 4X, 6 housing around electromechanical components.
Protection
Sample Cooling Industrial Grade refrigeration unit.
Temperature 0°C to 50°C (32°F to 122°F) without optional enclosure and heater.
Limits
Sample Pump High Speed peristaltic, dual roller design with impact and corrosion resistant Delrin
Safety Integral safety kill switch. E nsures when t he clear plastic face plate is removed from
®
plastic pump body.
the pump, it prevents accidental powered rotation.
Pump Tubing 3/8" ID by 5/8" OD by 1/8" thick medical grade silicone rubber pump tubing.
Tube Life Recommended maximum of 1,000,000 pump revolutions based on a standard sample.
A standard sample equates to 5ft of head, 10 foot PVC intake tube, and 200 ml sample
size.
Maximum Lift 28 ft (8.5344 m).
Transport Minimum of 3 ft/s at 3 ft of lift (0.9144 m/s at 1 m) and 2.0 ft/s at 20 ft of lift (0.6096
Velocity m/s at 6.1 m).
Sample Volume Programmed directly in increments of 1 milliliter up t o a maximum of 9,999 ml.
Accuracy ± 10ml or ± 10% of t he programmed volume, whichever is greater.
Repeatability ± 5ml o r ± 5% of t he average largest and smallest sample volume in a sample set,
whichever is greater.
Liquid Sensor Continuity type or Ultrasonic (optional)

Table of Contents