MODEL YB8 PROGRAMMING SECTION
April 1999/Manning Environmental Inc.
3DJH%
Display on LCD Explanation
SAMPLER READY
04/30 04:30:02
This display indicates the sampler is ready t o program and displays
the current time. Press the * key t o begin programming.
ENTER * MODE
__ __ 04:30:02
The sampler is now prompting fo r a star mode to be input. E nter the
numbers which represent the star mode of choice, in this example 09,
and press <ENTER>.
UPPER LEVEL LIMIT?
__ __ __ __ 04:30:02
Enter t he Upper Level Limit as a 4-digit number. R emember t he
decimal is implied in this program and the unit of measure is generic,
so if t he entry were t o be 10 feet/meters/inches/millimeters, it could
be entered as 0010, 0100, or 1000. Be CONSISTENT in all entries
throughout the program.
LOWER LEVEL LIMIT?
__ __ __ __ 04:30:02
Enter t he Lower Level Limit as a 4-digit number. R emember t o be
consistent with t he implied decimal from previous entries.
RISING DELTA?
__ __ __ __ 04:30:02
Enter a 4-digit number which represents the rising change in water
level that will t rigger a sample to be taken. I f t he rise of t he water is
equal t o o r greater t han this number a sample will be t aken.
FALLING DELTA?
__ __ __ __ 04:30:02
Enter a 4-digit number which represents the falling change in water
level that will t rigger a sample to be taken. I f t he fall o f the water is
equal t o o r greater t han this number a sample will be t aken.
SAMPLING LEVEL 1?
__ __ __ __ 04:30:02
Enter a 4-digit number that represents the lowest level at which a
sample is t o be taken. M ust be greater t han the Lower Level Limit.
Remember t o be consistent with t he implied decimal from previous
entries.
TIME OVERRIDE 1?
__ __ : __ __ 04:30:02
Enter a time in HH:MM format. T his is the amount of time after
Sampling Level 1 during which the sampler waits for an event
(Rising or Falling Delta, Sampling Level 2 reached, etc.). If no event
occurs before the interval is done, a sample will be t aken. I f an event
occurs, the Time Override will reset, or mo ve t o Time Override 2 if
Sampling Level 2 has been reached.
SAMPLING LEVEL 2?
__ __ __ __ 04:30:02
Enter a 4-digit number representing the next level at which a sample
should be taken, keeping consistent with the implied decimal point in
previous entries. A sample will be t aken when the water level rises to
this point. M ust be greater t han Sampling Level 1.
TIME OVERRIDE 2?
__ __ __ __ 04:30:02
Enter a time in HH:MM format. Operates on the same principal as
Time Override 1.