EasyManua.ls Logo

Manning YB8 - Page 82

Default Icon
121 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PROGRAMMING SECTION MODEL YB8
April 1999/Manning Environmental Inc.
3DJH%
SAMPLER READY
04/30 04:30:02
This display indicates the sampler is ready t o program and displays
the current time. Press the * key t o begin programming.
ENTER * MODE
__ __ 04:30:02
The sampler is now prompting fo r a star mode to be input. E nter the
numbers which represent the star mode of choice, in this example 11,
and press <ENTER>.
MAXIMUM FLOW RATE?
__ __ __ __ 04:30:02
Input the 4 most significant digits of the Maximum anticipated flow
rate. Since the unit of measurement is generic it can stand for any
volume/unit of time. If t he flo w rate is 40, it could be entered as:
4000, 0400, or 0040. The decimal point is implied, in each case, so
be consistent with all entries
. Rate must be in units per minute.
MINIMUM FLOW RATE?
__ __ __ __ 04:30:02
Enter the 4 most significant digits of the Minimum anticipated flow
rate. T he same criteria apply t o this input as t o Maximum Flow
Rate.
FLOW MULTIPLIER?
__ __ __ __ 04:30:02
The Flow Multiplier is used to scale the Maximum & Minimum Flow
Rates. If the Max flow rate is 40,000, enter it as 4000 (first 4
significant digits). The user would then enter a Flow Multiplier of 10
(4000 x 10 = 40,000) to have the unit scale the flow rate as 40,000.
SAMPLE TRIGGER?
__ __ __ __ 04:30:02
Enter the 4 most significant digits of t otalized flow at which a sample
should be taken. R emember the decimal point is implied and must be
consistent with previous entries. If the user wanted to take a sample
every 150,000 units, the entry would be 1500.
TRIGGER MULTIPLIER?
__ __ __ __ 04:30:02
The Trigger Multiplier scales the Sample Trigger. U sing t he example
above, if 150,000 units is the Trigger point, the Trigger Multiplier
would be 100 (1500 x 100 = 150,000).
SAMPLES PER BOTTLE?
__ __ 04:30:02
Enter the number of samples per bottle as a 2-digit number (1-99).
Make sure the volume to be placed in the bottles is not greater than
the actual volume of the bottles.
TIME OVERRIDE?
__ __ : __ __ 04:30:02
Enter in HH:MM format. R emember to allow enough t ime for t he
sampler to collect t he required samples.
PUSH START/OPTIONS
04:30:02
If no add-on options are desired, press <START> to begin the
program.
FLOW MODE (*11)
__ __ __ __ 04:30:02
The sampler will immediately begin reading t he analog signal.
Display on LCD Explanation

Table of Contents