MODEL YB8 PROGRAMMING SECTION
April 1999/Manning Environmental Inc.
3DJH%
risk that t he sampler will potentially not scale the analog signal correctly and will subsequently not t ake
samples at the anticipated or correct instances.
Display on LCD Explanation
SAMPLER READY
04/30 04:30:02
This display indicates the sampler is ready t o program and displays
the current time. Press the * key t o begin programming.
ENTER * MODE
__ __ 04:30:02
The sampler is now prompting fo r a star mode to be input. E nter the
numbers which represent the star mode of choice, in this example 13,
and press <ENTER>.
MAXIMUM FLOW RATE?
__ __ __ __ 04:30:02
Input the 4 most significant digits of the Maximum anticipated flow
rate. Since the unit of measurement is generic it can stand for any
volume/unit of time. If t he flo w rate is 40, it could be entered as:
4000, 0400, or 0040. The decimal point is implied, in each case, so
be consistent with all entries. Rate must be in units per minute.
MINIMUM FLOW RATE?
__ __ __ __ 04:30:02
Enter the 4 most significant digits of the Minimum anticipated flow
rate. T he same criteria apply t o this input as t o Maximum Flow
Rate.
FLOW MULTIPLIER?
__ __ __ __ 04:30:02
The Flow Multiplier is used to scale the Maximum & Minimum Flow
Rates. If the Max flow rate is 40,000, enter it as 4000 (first 4
significant digits). The user would then enter a Flow Multiplier of 10
(4000 x 10 = 40,000) to have the unit scale the flow rate as 40,000.
SAMPLE TRIGGER?
__ __ __ __ 04:30:02
Enter the 4 most significant digits of t otalized flow at which a sample
should be taken. R emember the decimal point is implied and must be
consistent with previous entries. If the user wanted to take a sample
every 150,000 units, the entry would be 1500.
TRIGGER MULTIPLIER?
__ __ __ __ 04:30:02
The Trigger Multiplier scales the Sample Trigger. U sing t he example
above, if 150,000 units is the Trigger point, the Trigger Multiplier
would be 100 (1500 x 100 = 150,000).
SAMPLES PER BOTTLE?
__ __ __ __ 04:30:02
Enter the number of samples per bottle as a 2-digit number (1-99).
Make sure the volume to be placed in the bottles is not greater than
the actual volume of the bottles.
TIME OVERRIDE?
__ __ : __ __ 04:30:02
Enter in HH:MM format. R emember to allow enough t ime for t he
sampler to collect t he required samples.
ENTER DELAY START
__ __ : __ __ 04:30:02
This display prompts the user to enter a delay start time (HH:MM
format). T his is the amount o f t ime the sampler is t o wait before
starting.