108 CME A required command parameter was missing.
109 CME An invalid TC_MEAS or RTD_MEAS unit parameter (not CEL
or FAR) was received, or an invalid PRES_UNIT or
ISO_PRES_UNIT parameter was received.
110 CME An invalid RANGELCK parameter was received.
111 EXE RANGELCK ON was received when the 3001 is not in Volts
mode.
112 CME An invalid RTD_TYPE parameter was received.
113 CME An invalid TC_REF parameter was received.
114 CME An invalid TSENS_TYPE parameter was received.
116 EXE No pressure module was present when a pressure function
was requested, or the total zero offset from calibration is more
than 6%. This error can also occur if the total zero offset from
calibration is out of limits when zeroing the thermocouple mil-
livolts type (maximum offset ±1 mV), or the RTD ohms type
(maximum ±0.1 ohm on high range, or ±0.01 ohm on the low
range).
117 CME An unrecognizable command was received.
118 CME An invalid parameter was received.
120 EXE The serial input buffer overflowed
121 EXE The command string buffer overflowed.
122 QYE The serial output buffer overflowed.
123 DDE The output overloaded. See display error message OVER
LOAD in section 2.7.
124 DDE The 3001 is out of tolerance. This error is set after a failed ini-
tialization or a failed *TST? command.
125 DDE The 3001 ADC has failed. This error is set after a failed initial-
ization or a failed *TST? command.
53