12. Specifications
12.1 General Specifications
Warm up time Twice the time since last warmed up, to a maximum of 30 minutes.
Settling time Less than 5 seconds for all functions and ranges except as noted.
Standard interfaces RS-232
IEEE-488 (GPIB)
Temperature performance Operating 0 °C to 50 °C
Calibration (tcal) 18 °C to 28 °C
Storage -20 °C to 70 °C
Electromagnetic compatibility CE: Conforms to EN61326
Temperature coefficient Temperature coefficient for temperatures outside tcal ±5 °C is 10% of
the 90 day specification (or 1 year if applicable) per °C
Relative humidity Operating <80% to 30 °C
<70% to 40 °C
<40% to 50 °C
Storage <95 % noncondensing
Altitude Operating 3,050 m (10,000 ft) maximum
Nonoperating 12,200 m (40,000 ft) maximum
Safety EN 61010 Second, ANSI/ISA-S82.01-1994,
CAN/CSA-C22.2 No. 1010.1-92, NRTL
Analog low isolation 20 V
Line power Line Voltage (selectable) 100V/120V or 220V/240V
Line Frequency 47 to 63 Hz
Line Voltage Variation ±10 % about setting
Power consumption 15 VA maximum
Dimensions Height 13.3 cm (5.25 in) plus 2.9 cm (1.15 in) for extended feet
Width standard rack width (19 inch)
Depth 30.0 cm (11.81 in) overall
Weight (without options) 4 kg (9 lb)
79