FUNC? IEEE-488 RS-232 Sequential Overlapped
This command returns the present output, measurement, or calibration function for the
primary and isolated displays.
Parameter: <None>
Response: <isolated>,<primary>
where <isolated> is one of the following:
DC10V measure DC voltage, 10V range
DC100V measure DC voltage, 100V range
DCI measure DC current
PRESSURE measure pressure
and where <primary> is one of the following:
DCV source DC voltage
DCI source DC current
RTD_OUT source RTD/Ohms
RTD_IN measure RTD/Ohms
TC_OUT source thermocouple
TC_IN measure thermocouple
PRESSURE measure pressure
Example: FUNC?
DC10V,PRESSURE
This example indicates that the isolated display is selected to the 10V
range and the primary display is selected to pressure.
HART? IEEE-488 RS-232 Sequential Overlapped
This command returns the isolated DC current input HART resistor status.
Parameter: <None>
Response: <value>
where <value> is ON or OFF
Example: HART?
OFF
This example indicates that the isolated DC current input HART resistor
is turned off.