ISO_MEAS IEEE-488 RS-232 Sequential Overlapped
This command sets the isolated measurement type.
Parameter: <value>
where <value> is one of the following:
DC10V measure DC voltage, 10V range
DC100V measure DC voltage, 100V range
DCI measure DC current
PRESSURE measure pressure
Response: <None>
Example: ISO_MEAS DCI
This example sets the isolated measurement to DC current.
ISO_PRES_UNIT IEEE-488 RS-232 Sequential Overlapped
This command sets the isolated pressure unit.
Parameter: <value>
where <value> is one of the following:
PSI pounds per square inch
INH2O4C inches of water at 4 °C
INH2O20C inches of water at 20 °C
INH2O60F inches of water at 60 °F
CMH2O4C centimeters of water at 4 °C
CMH2O20C centimeters of water at 20 °C
MMH2O4C millimeters of water at 4 °C
MMH2O20C millimeters of water at 20 °C
BAR bars
MBAR millibars
KPA kilopascals
MPA megapascals
INHG inches of mercury at 0 °C
MMHG millimeters of mercury at 0 °C
KG/CM2 kilograms per square centimeter
Response: <None>
Example: ISO_PRES_UNIT BAR
This example sets the isolated pressure unit to bars.