RANGELCK? IEEE-488 RS-232 Sequential Overlapped
This command returns the DC voltage range lock status.
Parameter: <None>
Response: <value>
where <value> is one of the following:
ON DC voltage range lock is on
OFF DC voltage range lock is off
Example: RANGELCK?
OFF
This example indicates that the range lock is off.
REMOTE IEEE-488 RS-232 Sequential Overlapped
This command places the 3001 into the remote state. It duplicates the IEEE-488 REN
(Remote Enable) message. When the 3001 is in the remote state, but not locked out,
only the LOCAL key is active. If the front panel is also locked out, no front panel keys are
active; see the LOCKOUT command. To unlock the front panel, use the LOCAL
command, or cycle the 3001 power switch.
Parameter: <None>
Response: <None>
Example: REMOTE
This example places the 3001 into the remote state.
*RST IEEE-488 RS-232 Sequential Overlapped
This command resets the 3001 to the power-up state and holds off execution of
subsequent commands until the reset operation is complete.
A reset action invokes the following commands and values for the primary display:
Command
Value
OUT 0 V
PRES_UNIT Last selected
RANGE 0.1 V
RTD_TYPE Last selected
STBY (No output)
TC_REF INT
TC_TYPE Last selected
TSENS_TYPE Last selected
The isolated display and selections remain as they were last selected.
Parameter: <None>
Response: <None>
Example: *RST
This example resets the 3001, invoking the commands and values
shown above.