STBY IEEE-488 RS-232 Sequential Overlapped
This command places the 3001 in standby mode, deactivating the output at front panel
terminals. This command acts the same as pressing the front panel key when in
operate mode.
Parameter: <None>
Response: <None>
Example: STBY
This example disconnects the selected output from the 3001 front
panel terminals. It also indicates Stby on the display.
TC_MEAS IEEE-488 RS-232 Sequential Overlapped
This command places the primary display in thermocouple measure mode.
Parameter: <value>
where <value> is one of the following:
CEL display in degrees celsius
FAR display in degrees fahrenheit
<None> display in the last selected temperature unit
Response: <None>
Example: TC_MEAS FAR
This example sets the 3001 to thermocouple measure mode,
displaying in degrees fahrenheit.
TC_REF IEEE-488 RS-232 Sequential Overlapped
This command selects the internal temperature sensor or an external reference value for
cold junction compensation of thermocouple source and measurement.
Parameter: <value>
where <value> is one of the following:
INT use internal temperature sensor
EXT use external reference value
Response: <None>
Example: TC_REF EXT
This example sets the thermocouple reference to external.