Thermocouple Specification, Output and Input (continued)
Absolute Uncertainty,
tcal ±5 °C, ±(°C)
1,2
Range (°C) Output/Input
TC Type Minimum Maximum 90 days 1 year
XK -200 °C -100 °C 0.22 °C 0.22 °C
-100 °C 300 °C 0.12 °C 0.13 °C
300 °C 800 °C 0.19 °C 0.20 °C
BP 0 °C 200 °C 0.42 °C 0.42 °C
200 °C 600 °C 0.32 °C 0.32 °C
600 °C 800 °C 0.39 °C 0.40 °C
800 °C 1600 °C 0.45 °C 0.46 °C
1600 °C 2000 °C 0.57 °C 0.58 °C
2000 °C 2500 °C 0.67 °C 0.80 °C
1. Does not include thermocouple wire error.
2. Absolute Uncertainty is inclusive of any cold juction compensation error, which is not stated
separately.
83