www.masimo.com 10 Masimo
WARNING: Do not use EMMA with nebulized medications as this may affect the light transmission of the EMMA
Airway Adapter windows.
WARNING: Audible alarm of any monitor may not be heard in some loud environments, such as when sirens are in use
and the care provider is more distant from the alarm source. Alarm volume should be tested with the extremes of your
noise environment to confirm ability or limitations to hear an alarm in all circumstances of the environment.
WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could
result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to
verify that they are operating properly.
WARNING: Make sure that EMMA is used in the electromagnetic environment specified in this manual.
WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the EMMA. Otherwise, degradation of the
performance of the EMMA could result.
Note: A trained medical professional must determine the proper EMMA Airway Adapter model for each patient
application. No hardware or software configuration changes result from the EMMA Airway Adapter model selected.
Note: The alarm limits will be reset to default values after EMMA powers off.
Cleaning and Service Warnings and Cautions
WARNING: Properly use and dispose of batteries or they may leak or explode.
WARNING: Lithium batteries may present a fire or chemical burn hazard if mistreated. Do not disassemble, heat
above 100°C (212°F) or incinerate. Dispose of used cells promptly. Keep away from children.
WARNING: Use only Alkaline batteries or Energizer Ultimate Lithium L92 batteries. Use of other Lithium batteries
may present a risk of fire or explosion.
WARNING: Replace batteries immediately when the Battery Status Indicator starts blinking. Remaining battery time
depends on battery type and other circumstances and cannot be reliably predicted.
CAUTION: Remove alkaline batteries when the EMMA will not be in use for more than 30 days to avoid damage to the
device due to batteries that may leak.