EasyManua.ls Logo

Masimo EMMA - Additional Gas Analyzer Specifications

Masimo EMMA
66 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
www.masimo.com 45 Masimo
Equipment Classification per IEC 60601-1
Degree of Protection of Electrical Shock
Defibrillation proof BF-Applied Part
Protection against harm from solid and
liquid ingress
IP44, Protection from tools and small wires greater than 1mm and from
water spray from any direction.
Mode of Operation Continuous operation
Sterility No part of EMMA is sterile
Additional Specifications
General Specifications
Description Compact, battery powered, quantitative capnograph for mainstream CO
2
monitoring
of adult, pediatric and infant patients.
Measurements [1] The CO
2
partial pressure is measured based on a 2 channel NDIR type gas analyzer at
4–5 μm with data acquisition rate at 10 kHz (sample rate 20 Hz / channel).
Models CO
2
displayed in kPa or mmHg

Other manuals for Masimo EMMA

Related product manuals