The audible alarm sound may be checked by detaching the EMMA Airway Adapter to generate a No Adapter alarm.
If the EtCO
2
Value is non-zero, ensure that there has not been an accumulation of CO
2
between the EMMA Sensor Body
and the EMMA Airway Adapter by removing and reattaching the EMMA Airway Adapter. If the EtCO
2
Value still
displays a non-zero value after this procedure, perform a Zeroing procedure. See Zeroing Procedure on page 56.
Power Off
The EMMA Capnograph switches off automatically during following conditions:
• If no breath is detected within 2 minutes from power up.
• If no breath is detected for 2 minutes and the alarm has been silenced.
• 15 seconds after the EMMA Airway Adapter is removed.