Chapter 7: Specifications
Display Range
Measurement Display Range
EtCO
2
(End-Tidal CO
2
) [1], [2] * 0 mmHg to 99 mmHg
0 kPa to 9.9 kPa
RR (Respiration Rate) 3 BPM to 150 BPM
* ETCO2 will be within specification for respiration rates up to 150 bpm [3]
Accuracy (ARMS)
Carbon Dioxide (CO
2
) [4]
Range 0 mmHg to 99 mmHg 0-40 mmHg ± 2 mmHg, 41-99 mmHg 6% of reading during standard conditions
Range 0 kPa to 9.9 kPa 0-5.3 kPa ± 0.3 kPa, 5.4-9.9 kPa 6% of reading during standard conditions

Other manuals for Masimo EMMA