www.masimo.com 35 Masimo
Chapter 5: Alarms and Messages
Alarms and Messages
The EMMA Capnograph is equipped with an Alarm Status Indicator and an audible alarm. The audible alarm can be
silenced for 2 minutes by pressing the Alarm Silence button. See Controls on page 29.
When the audible alarm is silenced, the yellow silence alarm indicator in the
bottom right corner of the display, i.e. the Alarm Status Silence Indicator, will
be lit.
Pressing the Alarm Silence button again during the 2 minutes mute period will
reactivate the audible alarm.
If a No Breath alarm is muted by pressing the Alarm Silence button, the EMMA
Capnograph will automatically switch off after 2 minutes provided that no new
breaths are detected.
If the alarm situation passes while the audible alarm is silenced, the alarm
icon turns green.
Pressing the Alarm Silence button during no alarm will also show a green
silence alarm indicator in the bottom right corner of the display.