www.masimo.com 39 Masimo
Chapter 6: Troubleshooting
Troubleshooting EMMA
Error Possible Causes Recommended Solutions
No Adapter alarm is
displayed
Indicates that an EMMA Airway
Adapter needs to be installed.
Connect an EMMA Airway Adapter. See Attaching
Airway Adapter on page 20.
Clogged Adapter alarm is
displayed
Indicates that the EMMA Airway
Adapter needs to be replaced with a
new one.
Replace the EMMA Airway Adapter with a new one.
Zero point adjustment
alarm is displayed
A zero point adjustment is required. Perform a Zeroing procedure. See Zeroing
Procedure on page 56.
The unit does not
complete the turn on
sequence
Low battery Replace the batteries.
The unit does not turn on
No battery
Low battery
Replace the batteries.

Other manuals for Masimo EMMA

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Masimo EMMA and is the answer not in the manual?

Masimo EMMA Specifications

General IconGeneral
ParametersEtCO2, Respiration Rate
Measurement Range0-150 mmHg
DisplayLCD
Battery LifeUp to 4 hours of continuous operation
Weight200 grams
TechnologyInfrared spectroscopy
EtCO2 Range0 to 150 mmHg
Respiration Rate Range0-99 breaths/min