Alarms and Messages
Active alarms are further displayed according to the following table:
Alarm Screen ETCO2 Value RR Value
No Breath NORMAL value steady
1
"- -" flashing
2
Low EtCO
2
NORMAL value flashing value steady
High EtCO
2
NORMAL value flashing value steady
Clogged Adapter ADAPTER n/a n/a
No Adapter ADAPTER n/a n/a
Zero point adjustment
3
NORMAL value steady value steady
Note 1: EtCO
2
value displays momentary CO
2
during No Breath.
Note 2: RR value displays "- -" steady if no breath at all detected from power on.
Note 3: Perform Zeroing procedure. See Zeroing Procedure on page 56.