Parameter Settings
The following information contains default alarm limits and information for adjusting EtCO
2
alarm limits.
Default Alarm Limits
The default factory settings for the RR and the EtCO
2
alarms are as follows:
Lower Limit Upper Limit
RR (No Breath) 3 bpm (20 seconds) NA
EtCO
2
OFF 50 mmHg (7.0 kPa)
EtCO2 Settings
The adjustment ranges for the EtCO
2
alarm limits are as follows:
Lower range Upper range
EtCO
2
(mmHg) OFF; 1 89 mmHg
11 99 mmHg; OFF
EtCO
2
(kPa) OFF; 0.1 8.9 kPa
1.1 9.9 kPa; OFF