www.masimo.com 58 Masimo
Barometric pressure
[mmHg]
EMMA Capnograph EtCO
2
readings should be between
5% CO
2
[mmHg] 5% CO
2
[kPa]
660-679 31-36 4,1-4,8
680-699 32-37 4,3-4,9
700-719 33-38 4,4-5,1
720-739 34-39 4,5-5,2
740-759 35-40 4,6-5,4
760-779 36-41 4,8-5,5
780-799 37-42 4,9-5,6
If the unit is reading within the above range then your EMMA Capnograph has been successfully verified.
If the unit is not reading within the above range, disconnect the EMMA Airway Adapter from the gas cylinder and
perform a Zeroing procedure and then repeat the Gas span check procedure. See Zeroing Procedure on page 56. If
verification still fails, contact your local distributor for further instructions.