Page #117 background imageLoading...
Page #117 background image
117
6688F-eIFU-0818
PERFORMANCE SPECIFICATIONS
The table below shows Arms (Accuracy Root Mean Square) values measured using the M-LNCS/LNCS Sensor with Masimo
SET Oximetry Technology in a clinical study.
M-LNCS/LNCS - Adtx/Pdtx
Range Arms
90-100% 1.64%
80-90% 1.07%
70-80% 1.55%
M-LNCS/LNCS - Inf/Neo
Range Arms
90-100% 1.85%
80-90% 1.44%
70-80% 0.89%
SaO2 versus error (SpO2 – SaO2) with linear regression fit and upper 95% and lower 95% limits of agreement
M-LNCS/LNCS - Adtx/Pdtx
Adt Masimo Saturation Accuracy
-10.0
-5.0
0.0
5.0
10.0
70 75 80 85 90 95
100
Error (SpO2-SaO2)
M-LNCS/LNCS - Inf/Neo/NeoPt
Neo Masimo Saturation Accuracy
-10.0
-5.0
0.0
5.0
10.0
70
75
80
85
90
95 100
SaO2
Error (SpO2-SaO2)