142
9123D-eIFU-0917
PERFORMANCE SPECIFICATIONS
Table information provides Arms values measured using the M-LNCS/LNCS/LNOP YI sensors with Masimo SET Oximetry Technology in a clinical
study.
MEASURED ARMS VALUES
RANGE
Arms
90-100% 0.6%
80-90% 0.5%
70-80% 0.7%
OVERALL CLAIMED ACCURACY VALUE
70-100% 2 %
SaO
2 versus error (SpO2 - SaO2) with linear regression fit and upper 95% and lower 95% limits of agreement.
M-LNCS/LNCS/LNOP YI
DC-I Saturation Accuracy
-10.0
-5.0
0.0
5.0
10.0
70 75 80 85 90 95 100
SaO2
Error (SpO2-SaO2)