Root Chapter 10- Specifications
www.masimo.com 89
Masimo
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines.
IEC 61000-4-11
100% for 0.5
cycle
60% for 5
cycles
30% for 25
cycles
100% for 5
seconds
-- Mains power quality should be
that of a typical commercial or
hospital environment. Root
provides a battery for continued
operation during power mains
interruption for a maximum of 4
hours.
Power frequency
(50 / 60 Hz)
magnetic field.
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields
should be at levels characteristic
of typical location in a typical
hospital environment.
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
3 Vrms

Table of Contents