During immersion in the enzymatic
solution, delicately brush the surface of the
wrench with a brush with soft nylon bristles
to eliminate all traces of visible dirt, both in
the internal part and the external one;
Delicately brush the surface of the wrench
with a brush with soft nylon bristles under
tap water;
3
4
5
6
7
Place the insert in the ultrasound tank
submerged by the enzymatic detergent
solution at 40°C, for at least 10 minutes;
Rinse with tap water to eliminate all
residues;
Rinse the wrench with distilled water until
it is clearly visible that the contaminants
have all been removed;