CLEANING OF THE CONSOLE AND OF THE FOOT PEDAL
MATERIALS NECESSARY
- Clean, soft cloths with low
ÀEHUUHOHDVH
- Detergent solution (pH 6-9) and, if need be,
non-aggressive disinfectant solution with
neutral pH (pH7)
CAUTION: If you wish to disinfect,
we recommend that you use water-based
disinfectant solutions with neutral pH
(pH7). Alcohol-based disinfectant solutions
and hydrogen peroxide are contra-
indicated, because they can fade the color
and/or damage the plastic materials. This
also holds true for chemical products such
as acetone and alcohol.
CLEANING METHOD
pH
6-9
05
CLEANING OF THE CONSOLE
AND OF THE FOOT PEDAL
1
PREPARATION
- Make sure that all the following accessories
have been removed/disconnected from the
device console (see Chapter 03):
electrical power supply cable
foot pedal
peristaltic pump tube
handpiece
KDQGSLHFHÀ[HGVXSSRUW
irrigation bag support rod
protective foil (if present)
supplementary grounding connector
(if present)
WARNING: Do not sterilize the device
console and the foot pedal: it could stop
working and cause harm to person and/or
damages to things.
WARNING: Always turn the device off
via the switch and disconnect it from the
electrical network before performing the
cleaning and disinfection interventions.
WARNING:The device is not protected
against penetration of liquids. Do not
spray liquids directly onto the surface of
the device.
CAUTION: Do not clean the device
console and the foot pedal under running
tap water.
CAUTION: Do not immerge the device
console and the foot pedal in liquids and/
or solutions of various nature.
Clean the surface of the device with a
FOHDQVRIWFORWKZLWKORZÀEHUUHOHDVH
dampened with a detergent solution
(pH 6-9). If you wish to disinfect, use a
non-aggressive disinfectant solution with
neutral pH (pH7), following the instructions
provided by the manufacturer of the
solution;