4
EN
Prepare an enzymatic detergent solution
with pH 6-9, as per the instructions of the
manufacturer;
Immerge the end part of the peristaltic
pump tube or the tube of the bone
grafting kit, if used, in a container
containing the enzymatic detergent
solution prepared;
&ORVHWKHÁRZRIWKHDGMXVWHU
'LVFRQQHFWWKHÁRZDGMXVWHUIURPWKH
irrigation bag and from the peristaltic
pump tube or from the tube of the bone
grafting kit, if used.
'LVSRVHRIWKHÁRZDGMXVWHUFRUUHFWO\
following the norms in force locally on the
disposal of hospital wastes;
02 CLEAN/FLUSH FUNCTION
CAUTION: CleanÁush function.
The CLEAN/FLUSH function must be used
after every treatment, before starting the
cleaning and sterilization procedures.
CAUTION: If the tubes are not cleaned,
the crystallization of the salts may seriously
damage the device.
CAUTION: The handpiece and the cord
cannot be separated.
1
2
3
The CLEAN/FLUSH function (according to the device used) enables to run a cleaning cycle
on the irrigation circuit of the handpiece used during the treatment, by following the steps
described in this chapter.