12
EN
MATERIALS NECESSARY
- Enzymatic detergent with pH 6-9
- Detergent solution (pH 6-9) and, if need be,
non-aggressive disinfectant solution with
neutral pH (pH7)
- Water
- Container for the immersion in the
enzymatic solution
- Ultrasound tank
&OHDQVRIWFORWKVZLWKORZÀEHUUHOHDVH
- Brush with soft nylon bristles
- Syringe
- Distilled water
- Thermodisinfector
- Compressed air
- Medical lubricant
CAUTION: If you wish to disinfect,
we recommend that you use water-based
disinfectant solutions with a neutral
pH (pH7). Alcohol-based disinfectant
solutions and hydrogen peroxide are
contra-indicated, because they can fade
the color and/or damage the plastic
materials. This also holds true for chemical
products such as acetone and alcohol.
Always rinse with sterile water to preserve
the disinfection.
CAUTION: Do not use hydrogen
peroxide. If you wish to disinfect the
accessories, do not use hydrogen
peroxide, but only disinfectants with
neutral pH (pH7); always rinse with sterile
water to preserve the disinfection.
06 CLEANING OF THE ACCESSORIES
2
PREPARATION
- Perform the CLEAN/FLUSH function
(see Chapter 02)
- Disconnect the handpiece from the device
console
- Disconnect the peristaltic pump
tube from the connector of the handpiece
- Remove the tube from the peristaltic pump.
If the bone grafting kit has been used,
disconnect the tube-to-tube connector
from both tubes
- If present, disconnect the insert from the
handpiece with the torque wrench
- Unscrew the front terminal of the handpiece
- Disconnect the irrigation bag support rod
from the device console
Refer to Chapter 3 for the disconnection of all
accessories.
CAUTION: The handpiece and the
cord cannot be separated.
CAUTION: Disconnect the handpiece
from the device exclusively acting on the
connector.
CAUTION: Always disconnect
the insert from the handpiece before
proceeding to its cleaning and
sterilization.
CAUTION: Do not immerge the
handpiece in disinfectant solutions
or other liquids, because it could get
damaged.
CAUTION: Do not immerge the
handpiece in an ultrasonic tank.
Dry the device console and the foot pedal
with a clean, non-abrasive cloth with low
ÀEHUUHOHDVH