22
EN
08 DRYING AND LUBRICATION
2QFH\RX·YHÀQLVKHGWKHFOHDQLQJ
operations, check all the objects under an
adequate source of light, if need be also
with a magnifying glass, paying attention
to the details that could conceal dirt
residues (threading, cavities, grooves) and,
if need be, repeat the cleaning cycle by
following the manual/automated cycle as
described in Paragraph 06.1. Finally, check
the integrity of those parts and elements
that may have worn out during their use;
07
CLEANING CHECK
1
1
Thoroughly dry all parts of the handpiece
with cord, especially the electrical contacts,
blowing on them with compressed air;
CAUTION: The electrical contacts
of the handpiece cord connector must
be dry, before and after the sterilization
cycle, before connecting the cord to
the device. Always make sure that
the connector electrical contacts are
perfectly dry; if need be, dry them by
blowing compressed air.