23
EN
Lubricate with medical degree lubricant in
the indicated point;
Dry the torque wrench using a clean and
VRIWFORWKZLWKORZÀEHUUHOHDVH
08 DRYING AND LUBRICATION
Thouroughly dry the accessories on all
their surfaces, internal and external, using a
FOHDQDQGVRIWFORWKZLWKORZÀEHUUHOHDVH
or by blowing compressed air, depending
on the case.
5
4
2
3
CAUTION: Before starting the
sterilization cycle, make sure that the
insert is thoroughly dry, both externally
and internally. For this purpose, blow
compressed air both externally and
through the internal thru hole; this will
prevent the onset of stains and haloes
on the surface or rust inside the insert.
CAUTION: do not use lubricants
based on oil or silicon.