Delicately brush the surface of the insert
again with the brush with soft nylon
bristles;
Rinse the internal canal of the insert
with distilled water injected under pressure
(3,8 bar) for at least 10 seconds to
eliminate all residues;
Leave to soak for 10 minutes at 40°C ±2°C
.
This procedure reduces the quantity of
blood, protein, and mucous present in the
insert;
During immersion into the enzymatic
solution, delicately brush the surface of
the insert with a brush with soft nylon
bristles to eliminate all traces of visible dirt.
$FFXUDWHO\FOHDQWKHGLIÀFXOWDUHDVVXFKDV
sharp edges and especially the interstices
between the cutting cusps;
06 CLEANING OF THE ACCESSORIES