17
EN
INSERTS TORQUE WRENCH
15
17
18
19
16
Put the wrench to soak in the enzymatic
detergent solution for 10 minutes at 40°C
±2°C;
During immersion in the enzymatic
solution, delicately brush the surface of the
wrench with a brush with soft nylon bristles
to eliminate all traces of visible dirt, both in
the internal part and the external one;
Delicately brush the surface of the wrench
with a brush with soft nylon bristles under
tap water;
Rinse the wrench with distilled water until it
is clearly visible that the contaminants have
all been removed;
Place the insert in the ultrasound tank
submerged by the enzymatic detergent
solution at 40°C ±2°C, for at least 10
minutes;
06 CLEANING OF THE ACCESSORIES